Author Vishnyakov, E. A. ♦ Kolesnikov, A. O. ♦ Kuzin, A. A. ♦ Negrov, D. V. ♦ Ragozin, E. N. ♦ Sasorov, P. V. ♦ Shatokhin, A. N.
Source United States Department of Energy Office of Scientific and Technical Information
Content type Text
Language English
Subject Keyword INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY ♦ APERTURES ♦ CHARGE-COUPLED DEVICES ♦ DIFFRACTION ♦ GRATINGS ♦ INCIDENCE ANGLE ♦ LASER-PRODUCED PLASMA ♦ LAYERS ♦ MOLYBDENUM ♦ REFLECTION ♦ SILICON ♦ SOFT X RADIATION ♦ SPATIAL RESOLUTION ♦ SPHERICAL CONFIGURATION ♦ WAVELENGTHS ♦ X-RAY SPECTROMETERS
Abstract A broadband stigmatic (imaging) soft X-ray ( λ > 120 Å) spectrometer is experimentally realised. The optical configuration of the spectrometer comprises a plane grazing-incidence reflection grating with a spacing varying across its aperture according to a preassigned law [a so-called varied line-space (VLS) grating] and a broadband spherical normal-incidence mirror with an aperiodic Mo/Si multilayer structure. The average plate scale amounts to ∼5.5 Å mm{sup −1}. The radiation is recorded with a matrix CCD detector (2048 × 1024 pixels of size 13 μm). The line spectra of the multiply charged ions LiIII and FV–FVII excited in laser-produced plasma are recorded with a spatial resolution of ∼26 μm and a spectral resolving power R ≈ 500 is experimentally demonstrated. (paper)