Author Gargett, M. ♦ Petasecca, M. ♦ Alnaghy, S. ♦ Rosenfeld, A. ♦ Oborn, B. ♦ Metcalfe, P.
Source United States Department of Energy Office of Scientific and Technical Information
Content type Text
Language English
Subject Keyword APPLIED LIFE SCIENCES ♦ ACCURACY ♦ FILM DOSIMETRY ♦ LINEAR ACCELERATORS ♦ LUNGS ♦ PHANTOMS ♦ PHOTOGRAPHIC FILM DOSEMETERS ♦ READOUT SYSTEMS ♦ SI SEMICONDUCTOR DETECTORS
Abstract Purpose: To assess the capability of a novel 2D monolithic silicon detector array in measuring stereotactic photon fields. Methods: The silicon array detector used in this work, named Magic Plate-512 (MP512), is a thin monolithic silicon wafer (52 × 52 × 0.47 mm{sup 3}) with 512 ion-implanted diodes (0.5 × 0.5 mm{sup 2}). Adjacent pixels are spaced evenly with 2 mm pitch, covering a maximum detection area 46 mm wide. Its fast, FPGA based read-out system is synchronised with the linac to allow readout of all pixels pulse-by-pulse. A clinical SABR lung plan (consisting of 9 single segment beams, 6MV) was measured with the array at 1.5 cm depth in a solid water phantom (100 cm SSD). The typical field size was in the range of 3 × 3 cm{sup 2} to 4 × 4 cm{sup 2}. Each beam was delivered at perpendicular incidence to the detector plane so as to avoid the need for angular dependence corrections. The fields were measured under the same conditions using Gafchromic EBT3 film for comparison. The film was scanned at 72 dpi resolution, with the red channel data used for analysis. Results: Average gamma passing rates of (92.3 ± 1.8) % for 2%/2mm criteria, and (86.6 ± 2.3) % for 1%/2mm criteria were achieved for MP512, using EBT3 film as the reference distribution. The detector array was able to accurately measure the full-width-at-half-maximum (FWHM), to within (0.77 ± 0.01) mm accuracy when compared to film. The penumbral widths (80%-20%) were measured to within (0.30 ± 0.01) mm accuracy to film. Conclusion: The MP512 is a feasible option for measurement of stereotactic photon fields, with its high density of detection points making it useful for small field applications. The prototype array has demonstrated merit; in the future the development of a larger array detection area would be beneficial for clinical applications.
ISSN 00942405
Educational Use Research
Learning Resource Type Article
Publisher Date 2015-06-15
Publisher Place United States
Journal Medical Physics
Volume Number 42
Issue Number 6


Open content in new tab

   Open content in new tab