Thumbnail
Access Restriction
Open

Author Lombigit, L. ♦ Rahman, Nur Aira Abd ♦ Mohamad, Glam Hadzir Patai ♦ Ibrahim, Maslina Mohd ♦ Yussup, Nolida ♦ Yazid, Khairiah ♦ Jaafar, Zainudin
Source United States Department of Energy Office of Scientific and Technical Information
Content type Text
Language English
Subject Keyword CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS ♦ AMPLIFIERS ♦ GRIDS ♦ NOISE ♦ PULSES ♦ RADIOISOTOPES ♦ SENSITIVITY ♦ TESTING
Abstract A radioisotope identifier device based on large volume Co-planar grid CZT detector is current under development at Malaysian Nuclear Agency. This device is planned to be used for in-situ identification of radioisotopes based on their unique energies. This work reports on electronics testing performed on the front-end electronics (FEE) analog section comprising charge sensitive preamplifier-pulse shaping amplifier chain. This test involves measurement of charge sensitivity, pulse parameters and electronics noise. This report also present some preliminary results on the spectral measurement obtained from gamma emitting radioisotopes.
ISSN 0094243X
Educational Use Research
Learning Resource Type Article
Publisher Date 2016-01-22
Publisher Place United States
Volume Number 1704
Issue Number 1


Open content in new tab

   Open content in new tab