Subject Keyword INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY ♦ PLASMA PHYSICS AND FUSION TECHNOLOGY ♦ CONFINEMENT ♦ COUNTING RATES ♦ CRYSTALS ♦ DATA ACQUISITION ♦ DATA ACQUISITION SYSTEMS ♦ EQUIPMENT ♦ IMAGES ♦ KOREAN ORGANIZATIONS ♦ MULTIWIRE PROPORTIONAL CHAMBERS ♦ PHOTONS ♦ PLASMA ♦ SPECTROMETERS ♦ TOKAMAK DEVICES ♦ WIRES ♦ X RADIATION
Abstract The inconsistency of the first experimental results from the X-ray imaging crystal spectrometer for the Korea Superconducting Tokamak Advanced Research device utilizing a multi-wire proportional counter (MWPC) is clarified after improving the photon-count rate of the data acquisition system for the MWPC and ground loop isolator for the whole spectrometer system. The improved MWPC is successfully applied to pure Ohmic plasmas as well as plasmas with high confinement modes.