Author Li, Taibao
Source J-STAGE
Content type Text
Publisher Acoustical Society of Japan
Language English
Subject Keyword Pulsed Doppler ♦ Motion artifact reduction ♦ Implementation ♦ Phase compensation
Abstract A new technique to remove hand motion artifacts in pulsed Doppler is proposed. The instantaneous phase difference between two consective echoes is estimated and is supposed to be dominated by the hand motion component. The phase difference is then used to compensate the phase of the next echo. The same procedure is repeated in every transmit/receive cycle. Implementation of this technique is accomplished on a medical imaging system with a pulsed Doppler processor. The results observed shows that hand motion artifacts have been greatly reduced.
ISSN 03882861
Learning Resource Type Article
Publisher Date 1997-01-01
e-ISSN 21853509
Journal Journal of the Acoustical Society of Japan (E)(ast1980)
Volume Number 18
Issue Number 5
Page Count 5
Starting Page 267
Ending Page 271