Thumbnail
Access Restriction
Open

Author Subrahmanyam, K.
Source United States Department of Energy Office of Scientific and Technical Information
Content type Text
Language English
Subject Keyword INSTRUMENTS AND INSTRUMENTATION ♦ ELECTRON BEAMS ♦ ELECTRON TUBES ♦ ELECTRONIC EQUIPMENT ♦ RADIOISOTOPES ♦ RECORDING SYSTEMS ♦ SCINTILLATION COUNTERS ♦ SPECTRA ♦ SPECTROMETERS ♦ STABILITY
Abstract A set-up for quick qualitative and quantitative analysis of radioactive isotopes with a modified single channel scintillation spectrometer is described. The whole operation is automatic and the stability and the accuracy of the electronic equipment is found to be good when compared to conventional set-ups. Activities ranging between few microcuries to few milli microcuries are acceptable. The total time of operation can be varied from a few minutes to a few tens of minutes. The spectrum can either be displayed on a C.R. tube of sufficiently long persistance or can be recorded on a recorder. (auth)
Educational Use Research
Learning Resource Type Article
Publisher Date 1963-04-01
Journal Nucl. Instr. Methods
Volume Number 22
Organization Atomic Energy Establishment, Trombay, India


Open content in new tab

   Open content in new tab